Flow Sensor : ND Series

Feature

General

ND Flowsensor is ideal for measuring variety of fluid and outputs open collector pulses with the built-in amplifier type magnetic sensor.

Enhanced basic functions, user friendliness and cost efficiency are the key points

ND series offers high accuracy measurement of not only cold and warm water but also pure water, chemical fluids, etc. It is light-weighted compact body and is compatible with various kinds of fluids with the carefully selected casing materials such as PP, ETFE, PPO, etc., and with the highly sensitive magnetic sensor of our own development. This series is therefore widely applicable for various flow measurements and its control and monitoring as well.

Feature

  • Highly accurate over a wide measurement range
  • Simple and durable structure
  • Compact and lightweight (eg. ND10: 120g)
  • Compatible with various kinds of fluids
  • Free installation position (For ND and NW,except the position that  of the holding plate or the local display is faced downward.)
  • Excellent cost-performance
  • High sensitive high-density pulse output

Flowsensor Applications

  • To control of warm water temperature
  • To constantly supply water and warm water
  • To monitor coolant, etc.
  • To proportionally inject chemical fluids
  • To control pressure loss of filters

 

Overview

ItemModel
ND05 -
TATAAA

ND05-
PATAAC
ND05-
NATAAC
ND10-
TATAAA
ND10-
PATAAA
ND10-
NATAAA
ND20-
PATAAA
ND20-
NATAAA
Flow-rate range 0.3 - 3 L/min 1.0 - 10L/min 1.5 - 20L/min 3.0 - 60L/min
Accuracy
(Standard installation position)
±2% (RS)
Pulse unit 2.5mL/p 7.69mL/p 25mL/p
Pressure drop 12 kPa or less
(At 3L/min)
15 kPa or less
(At 10L/min)
20 kPa or less
(At 20L/min)
60 kPa or less
(At 60L/min)
Fluid temperature range 0 to +60°C 0 to +70°C 0 to +60°C 0 to +70°C 0 to +60°C 0 to +70°C
Connection R 1/2 R 3/4
Weight Approx.150g Approx.120g

Approx.360g

Common Specification

Maximum working pressure 1MPa or less with a fluid temperature of +20°C
Viscosity 0.5 to 1.5 mPa·s
Output signal Open collector pulse
4-wires, Lead-wire length approx. 600mm
Output signal
Duty ratio 3/7 < A/B < 7/3
Impressed voltage range 3 to 24 V DC
Ambient temperature -20° to +80° (Dew condensation must be avoided)
Measurable fluids Refer to “MEASURABLE FLUID AND PARTS MATERIALS”

 

Dimensions

 

 ND05ND10ND20
A 80 80 110
B 47 47 68
C 37.5 37.5 50
D 49 49 65
E 47 47 68
F 12.5 12.5 18
G 16 16 23

 

Handling Instructions

Observe the following handling instructions and precautions to obtain maximum performance of the Flowsensor / Flowmeter.

CAUTIONS IN MOUNTING

1. The flowing direction of the fluid, when mounting, must be matched to the arrow mark on the case that indicates the direction of flow.

2. A straight pipe with uniform inner diameter must be connected to the inlet. Accuracy is adversely affected if the pipe is bent or the inner diameter is uneven.

3. The inner diameter of the inlet pipe must be greater than the nozzle diameter of the Flowsensor / Flowmeter.

 

TypeInner D, of inlet pipeLength of straight pipe at inlet
ND05,NW05 5.5 mm or greater Greater than 50 mm from the end of the Flowsensor / Flowmeter
ND10,NW10 10.5 mm or greater Greater than 100 mm from the end of the Flowsensor / Flowmeter
ND20,NW20 15.5 mm or greater

 

4. If pulsation occurs in the flow of the fluid, measurement accuracy is adversely affected. When a pump is used to circulate the fluid, an accumulator must be provided to remove the pulsation.

5. Be careful not to apply excessive stress on the Flowsensor / Flowmeter during installation.

6. Make sure to provide thermal insulator at places subject to freezing in winter. When leakage occurs due to half frozen fluid, accuracy is adversely affected.

7. Do not expose/install the Flowsensor / Flowmeter under direct sunshine.

8. The measuring range, operating pressure and operating temperature are indicated on the plate on the side of the Flowsensor / Flowmeter. Observe these conditions for optimum performance of the Flowsensor / Flowmeter.

9. Avoid locating the Flowsensor / Flowmeter at places subject to excessive pressure and stress such as water hammer by operating components.

10. When air enters the Flowsensor / Flowmeter, measurement accuracy is influenced. Always operate the Flowsensor / Flowmeter chamber full of fluid. To maintain high accuracy, avoid mixture of air to the fluid.

11. Avoid placing a strong magnetic material or magnetic field near the Flowsensor / Flowmeter.

12. Installation position is flexible except those shown in figures on the upper right. However, it is recommended to install the Flowsensor / Flowmeter at the standard position whenever possible. For types ND 05-TATAAA, ND10-TATAAA, NW05-TTN, NW10-TTN, installation must be at standard positions.

 
Standard mounting position

Incorrect mounting position (the mark of or  the indicator is placed upside down)

 

CONNECTION (WIRING)

 

Measurable fluids and parts materials

MEASURABLE FLUIDS
  1. Fluid which done not corrode materials that contact the fluid.
  2. Viscosity of the fluid to be measured is between 0.5 and 1.5cp.
  3. No slurry is contained.
  4. Particles attracted by magnetic power such as iron, is contained in the fluid.
    (All of the above 1-4 must be satisfied.)
LIST OF MATERIALS THAT CONTACT FLUIDS DIRECTLY

The table below shows components and materials of the standard type Flowsensor / Flowmeter used in portions that come into direct contact with the fluid.

Parts nameModel
ND05-PATAAC
NW05-PTN
NW05-PTP
ND05-NATAAC
NW05-NTN
NW05-NTP
ND05-TATAAA
NW05-TTN
NW05-TTP
ND10-TATAAA
NW10-TTN
NW10-TTP
ND10-PATAAA
NW10-PTN
NW10-PTP
ND20-PATAAA
NW20-PTN
NW20-PTP
ND10-NATAAA
NW10-NTN
NW10-NTP
ND20-NATAAA
NW20-NTN
NW20-NTP
Lower casing
Upper casing
Regulator
PP Denatured PPO ETFE PP PP Denatured PPO
Vane-wheel POM POM POM POM POM
Pivot PA SUS304 PA SUS304 SUS304
O-ring FKM NBR FKM FKM FKM NBR
Magnet Samarium cobalt (No contact with fluid) Samarium cobalt (No contact with fluid) Samarium cobalt (No contact with fluid) Ba-Fe Ba-Fe Ba-Fe
Measurable fluid Water, 
Chemical fluid
(Weak corrosion resistance)
Water, 
Warm Water
Chemical fluid Water, Chemical Fluid
(Weak corrosion resistance)
Water, 
Warm Water
 
Material Code
  • Ba-Fe: Barium-Ferrite
  • ETFE: Ethylene-tetrafluoro ethylene
  • FKM: Fluoro Rubber
  • NBR: Acrylonitrile-butadiene rubber
  • PA: Polyamide
  • POM: Polyacetal or Polyoxymethylene
  • PP: Polypropylene
  • PPO: Polyphenylene oxide
  • SUS304: Stainless steel

 

 

 

Cr. : https://www.aichitokei.co.jp/eng/